{"sheets":[{"name":"Traceability","hidden":false,"max_col":"C","rows":{"1":{"A":"Lab ID","B":"Masters","C":"Traceability"},"2":{"A":"TEQ030","B":"Mass Flow meter, Make: Krohne, Model: IFS4000, Size: DN25, Serial No. G140000008305187, Range: 0 to 5 m³\/hr","C":"Traceable to National Standard through NABL Lab Certif. No CC-3526, Ahmedabad, Certificate No. IPC-11092024-047, Cal Date : 11\/09\/2024, Valid up to 10\/09\/2025"},"3":{"A":"TEQ031","B":"Magnetic Flow meter, Make : Krohne, Model : IFS4000, Size : DN100, Serial No. I14404227, Range : 0 to 250 m³\/hr","C":"Traceable to National Standard through NABL Lab Certif. No CC-3526, Ahmedabad, Certificate No. IPC-11092024-049, Cal Date : 11\/09\/2024, Valid up to 10\/09\/2025"},"4":{"A":"TEQ033","B":"Mass Flow meter, Make: Krohne, Model: IFS4000, Size: DN50, Serial No. G140000008505124, Range: 0 to 50 m³\/hr","C":"Traceable to National Standard through NABL Lab Certif. No CC-3526, Ahmedabad, Certificate No. IPC-11092024-048, Cal Date : 11\/09\/2024, Valid up to 10\/09\/2025"}}},{"name":"Data","hidden":false,"max_col":"N","rows":{"4":{"B":"Unit","C":"Conversation factor","D":"Roundup"},"5":{"B":"m³\/hr","C":"1","D":"0","G":"Range","H":"Unit","I":"Con. Factor"},"6":{"B":"m³\/min","C":"0.0166666667","D":"2","G":"9","H":"m³\/hr","I":"1","J":"m³\/hr","K":"1","L":"1"},"7":{"B":"m³\/s","C":"0.0002777778","D":"3","G":"9","H":"m³\/hr"},"8":{"B":"cm³\/hr","C":"1000000","D":"0","G":"2","H":"1","K":"0.0001","L":"4","M":"0","N":"4"},"9":{"B":"cm³\/min","C":"16666.6666666667","D":"0"},"10":{"B":"cm³\/s","C":"277.7777777778","D":"0"},"11":{"B":"ft³\/hr","C":"35.31466672","D":"0","F":"1","G":"TEQ030","H":"5","I":"0.0001","J":"1","K":"0.24"},"12":{"B":"ft³\/min","C":"0.5885777787","D":"0","F":"2","G":"TEQ033","H":"50","I":"0.0001","J":"1","K":"0.22"},"13":{"B":"ft³\/s","C":"0.0098096296","D":"2","F":"3","G":"TEQ031","H":"250","I":"0.001","J":"1","K":"0.22"},"14":{"B":"lps","C":"0.2777777778","D":"0","F":"4","G":"TEQ032","H":"705","I":"0.001"},"15":{"B":"LPM","C":"16.6666666667","D":"0"},"16":{"B":"lph","C":"1000","D":"-3"},"17":{"B":"ml\/h","C":"1000000","D":"0","G":"8.5","H":"2"},"18":{"B":"ml\/min","C":"16666.6666666667","D":"0","F":"2","G":"TEQ033"},"19":{"B":"ml\/s","C":"277.7777777778","D":"0","F":"2","G":"TEQ033"},"20":{"G":"E"}}},{"name":"Observation","hidden":false,"max_col":"AN","rows":{"2":{"E":"Observation Sheet for Flow Calibration"},"3":{"E":"Issue No","H":"01","L":"Issue Date","O":"01-01-2020"},"4":{"E":"Amend No","H":"00","L":"Amend Date","O":"-"},"6":{"A":"COP No","D":"Year","F":"Location","H":"SRF No","J":"Job No","M":"Calibration Date","P":"Certificate No"},"7":{"A":"COP\/005-01","D":"2025","F":"0","H":"LM251224\/1","J":"LM251224\/1 \/ TC \/ FL \/ 002","M":"24\/12\/2025","P":"LM251224\/1 \/ TC \/ FL \/ 002"},"8":{"A":"Basic Details of Device Under Calibration (DUC)"},"9":{"A":"Nomenclature of DUC","E":"Digital Flow meter (Electromagnetic)"},"10":{"A":"Range","E":"0.45","F":"to","G":"9","H":"m³\/hr","J":"Identification","N":"ID_19"},"11":{"A":"Resolution","E":"0.01","F":"m³\/hr","I":"2","J":"Serial No","N":"SR_19"},"12":{"A":"Make","E":"Accumax","J":"Model No","N":"AIPL_WM"},"13":{"A":"Calibration Point","E":"1.1 m³\/hr, 2.5 m³\/hr, 4 m³\/hr, 6 m³\/hr, 8.5 m³\/hr"},"14":{"A":"Size","E":"25 NB (1\")","J":"K Factor (Primary Constant)","N":"0.23335"},"15":{"A":"Environmental Condition"},"16":{"A":"Parameter","D":"Starting of Cal.","G":"Ending of Cal.","J":"Parameter","M":"Starting of Cal.","P":"Ending of Cal."},"17":{"A":"Temperature in °C","D":"28","G":"30","J":"Relative Humidity in %Rh","M":"32","P":"35","AD":"1.1","AE":"0.5"},"18":{"A":"Parameter","D":"Starting of Cal.","G":"Ending of Cal.","AD":"9","AE":"0.1"},"19":{"A":"Water Temperature in °C","D":"22","G":"23"},"20":{"A":"Master to be used from Following","G":"TEQ030"},"21":{"A":"Sr. No","B":"Description of master","G":"Identification No & Serial No","M":"Range \/ Resolution","R":"Tick","AD":"1.15","AE":"0.4974683544","AF":"0.4974683544","AG":"0.49","AH":"0.15"},"22":{"A":"1","B":"Mass Flow Meter","G":"Lab ID : TEQ030 \/ G140000008305187","M":"0 - 5 m³\/h(1.39 LPS) \/ 0.0001 LPS","AD":"2.555","AE":"0.1","AF":"0.4263291139","AG":"0.1","AH":"0.55"},"23":{"A":"2","B":"Mass Flow Meter","G":"Lab ID : TEQ033 \/G140000008505124","M":"0 - 50 m³\/h(13.9 LPS) \/ 0.001 LPS","AD":"4","AE":"0.1","AF":"0.353164557","AG":"0.1","AH":"0.04"},"24":{"A":"3","B":"Electromagnetic Flow Meter","G":"Lab ID : TEQ033 \/ I14404227","M":"0 - 250 m³\/h(69.45 LPS) \/ 0.001 LPS","S":"0.5","AD":"6","AE":"0.1","AF":"0.2518987342","AG":"0.1","AH":"0.06"},"26":{"A":"Sr. No.","B":"Calibration Point","C":"DUC Reading","D":"Reading Change","E":"Error (% fs)"},"27":{"A":"1","B":"1.1","C":"1.15","D":"0.48","E":"0.5","F":"3.781","G":"3.781"},"28":{"A":"2","B":"2.5","C":"2.555","D":"0.51","E":"0.1","F":"0.323","G":"0.323"},"29":{"A":"3","B":"4","C":"4","D":"0.53","E":"0.1","F":"0.231","G":"0.231"},"30":{"A":"4","B":"6","C":"6","D":"0.57","E":"0.1","F":"0.154","G":"0.154"},"31":{"A":"5","B":"8.5","C":"8.5","D":"0.62","E":"0.1","F":"0.108","G":"0.108"},"32":{"A":"Calibration Result","S":"0.1","AD":"8.5","AE":"0.1","AF":"0.1253164557","AG":"0.1","AH":"0.005"},"33":{"A":"1","B":"Error of Measurement","AD":"0","AG":"0.49"},"34":{"A":"Sr. No.","B":"Calibration Point in m³\/hr","D":"Reading on Master in lps","F":"Reading on DUC in m³\/hr","H":"Error in _____","J":"Sr. No.","K":"Calibration Point in _____","M":"Reading on Master in _____","O":"Reading on DUC in _____","Q":"Error in _____","AD":"1","AE":"9","AF":"4"},"35":{"B":"m³\/hr","D":"m³\/hr","F":"m³\/hr","H":"m³\/hr","K":"m³\/hr","M":"m³\/hr","O":"m³\/hr","Q":"m³\/hr","Y":"2"},"36":{"A":"1","B":"1.1","D":"1.1052","F":"1.15","H":"0.0448","J":"2","K":"2.5","M":"2.546","O":"2.555","Q":"0.009","AB":"0.0447721519","AD":"1.1052278481","AE":"1.15","AF":"0.4974683544","AH":"2.546","AI":"2.555","AJ":"0.1","AN":"1.1052"},"37":{"D":"1.1146","F":"1.16","H":"0.0454","M":"2.5309","O":"2.54","Q":"0.0091","AB":"0.045439257","AD":"1.114560743","AE":"1.16","AF":"0.5048806329","AH":"2.530901","AI":"2.54","AJ":"0.1011","AN":"1.1146"},"38":{"D":"1.1241","F":"1.17","H":"0.0459","M":"2.5608","O":"2.57","Q":"0.0092","AB":"0.0459390888","AD":"1.1240609112","AE":"1.17","AF":"0.5104343199","AH":"2.560800911","AI":"2.57","AJ":"0.1022121","AN":"1.1241"},"39":{"D":"1.1036","F":"1.15","H":"0.0464","M":"2.5407","O":"2.55","Q":"0.0093","AB":"0.0464444188","AD":"1.1035555812","AE":"1.15","AF":"0.5160490974","AH":"2.540699721","AI":"2.55","AJ":"0.1033364331","AN":"1.1036"},"40":{"D":"1.093","F":"1.14","H":"0.047","M":"2.5306","O":"2.54","Q":"0.0094","AB":"0.0469553074","AD":"1.0930446926","AE":"1.14","AF":"0.5217256375","AH":"2.5305611469","AI":"2.54","AJ":"0.104876146","AN":"1.093"},"41":{"B":"Average","D":"1.1081","F":"1.154","H":"0.0459","K":"Average","M":"2.5418","O":"2.551","Q":"0.0092"},"42":{"A":"3","B":"4","D":"3.991","F":"4","H":"0.009","J":"4","K":"6","M":"5.991","O":"6","Q":"0.009","AD":"3.991","AE":"4","AF":"0.1","AH":"5.991","AI":"6","AJ":"0.1","AN":"3.991"},"43":{"D":"3.9909","F":"4","H":"0.0091","M":"5.9909","O":"6","Q":"0.0091","AD":"3.990901","AE":"4","AF":"0.1011","AH":"5.990901","AI":"6","AJ":"0.1011","AN":"3.9909"},"44":{"D":"3.9908","F":"4","H":"0.0092","M":"5.9908","O":"6","Q":"0.0092","AD":"3.990800911","AE":"4","AF":"0.1022121","AH":"5.990800911","AI":"6","AJ":"0.1022121","AN":"3.9908"},"45":{"D":"3.9907","F":"4","H":"0.0093","M":"5.9907","O":"6","Q":"0.0093","AD":"3.990699721","AE":"4","AF":"0.1033364331","AH":"5.990699721","AI":"6","AJ":"0.1033364331","AN":"3.9907"},"46":{"D":"3.9906","F":"4","H":"0.0094","M":"5.9906","O":"6","Q":"0.0094","AD":"3.990597418","AE":"4","AF":"0.1044731339","AH":"5.990597418","AI":"6","AJ":"0.1044731339","AN":"3.9906"},"47":{"B":"Average","D":"3.9908","F":"4","H":"0.0092","K":"Average","M":"5.9908","O":"6","Q":"0.0092"},"48":{"A":"5","B":"8.5","D":"8.491","F":"8.5","H":"0.009","J":"6","Q":"0","AD":"8.491","AE":"8.5","AF":"0.1","AH":"-0.009","AI":"0","AJ":"0.1","AN":"8.491"},"49":{"D":"8.4909","F":"8.5","H":"0.0091","Q":"0","AD":"8.490901","AE":"8.5","AF":"0.1011","AH":"-0.0091341","AI":"0","AJ":"0.10149","AN":"8.4909"},"50":{"D":"8.4908","F":"8.5","H":"0.0092","Q":"0","AD":"8.490800911","AE":"8.5","AF":"0.1022121","AH":"-0.0092345751","AI":"0","AJ":"0.10260639","AN":"8.4908"},"51":{"D":"8.4907","F":"8.5","H":"0.0093","Q":"0","AD":"8.490699721","AE":"8.5","AF":"0.1033364331","AH":"-0.0093361554","AI":"0","AJ":"0.1037350603","AN":"8.4907"},"52":{"D":"8.4906","F":"8.5","H":"0.0094","Q":"0","AD":"8.4905611469","AE":"8.5","AF":"0.104876146","AH":"-0.0094388531","AI":"0","AJ":"0.104876146","AN":"8.4906"},"53":{"B":"Average","D":"8.4908","F":"8.5","H":"0.0092","K":"Average","Q":"0"},"55":{"A":"Unit Conversation with respect to SI Unit"},"56":{"A":"m³\/hr","F":"lps (Litre per second)","J":"LPM (Litre per Minute)","N":"LPH (Litre per Hour)"},"57":{"A":"1","F":"0.278","J":"16.667","N":"1000"}}},{"name":"Certificate","hidden":false,"max_col":"N","rows":{"2":{"A":"Certificate No","D":"LM251224\/1 \/ TC \/ FL \/ 002","G":"Unique Lab Report (ULR) ID","J":"CC234525000000206F"},"3":{"A":"Date of Calibration","D":"24\/12\/2025","G":"Recommended Next Due Date*","J":"24\/12\/2026"},"4":{"A":"Receipt \/ WRF No.","D":"LM251224\/1","G":"Date of DUC Received","J":"24\/12\/2025"},"5":{"A":"Certificate Issue Date","D":"25\/12\/2025","G":"Calibration Performed at","J":"Lab"},"6":{"A":"Details of Customer"},"7":{"A":"Name","D":"AIPL"},"8":{"A":"Address","D":"B-95, Sector-25 GIDC, Gandhinagar, Gujarat, India, 382016"},"9":{"A":"Ref. Doc. No. (GP\/PO)","D":"LTB-24122025-001"},"10":{"A":"Date of Doc.","D":"24\/12\/2025"},"11":{"A":"Details of Device under Calibration (DUC)"},"12":{"A":"Nomenclature","C":"Digital Flow meter (Electromagnetic)"},"13":{"A":"Range of Meter","C":"0.45 to 9 m³\/hr","G":"Mfg. Serial No","I":"SR_19"},"14":{"A":"Resolution","C":"0.01 m³\/hr","G":"Identification No","I":"ID_19"},"15":{"A":"Size","C":"25 NB (1\")","G":"Model No","I":"AIPL_WM"},"16":{"A":"Make","C":"Accumax","G":"Condition of DUC","I":"Good"},"17":{"A":"Specified Accuracy","C":"± 0.5% FS ( ±0.4 m³\/hr)","M":"E"},"18":{"A":"Details of Calibration Masters and Traceability"},"19":{"A":"Lab ID","B":"Description of Master","F":"Traceability"},"20":{"A":"TEQ030","B":"Mass Flow meter, Make: Krohne, Model: IFS4000, Size: DN25, Serial No. G140000008305187, Range: 0 to 5 m³\/hr","F":"Traceable to National Standard through NABL Lab Certif. No CC-3526, Ahmedabad, Certificate No. IPC-11092024-047, Cal Date : 11\/09\/2024, Valid up to 10\/09\/2025"},"21":{"A":"E","M":"E"},"22":{"A":"Details of Calibration Procedure and Reference Documents"},"23":{"A":"Lab Procedure","E":"Method of Calibration"},"24":{"A":"Lab Procedure No. : COP\/005-01","E":"The calibration of DUC was carried out by comparison with Standard flow meter by using Calibration Liquid as Water"},"25":{"A":"Calibration Method is generally based","E":"NABL 129-2019"},"26":{"A":"Details of Environment Condition at the Time of Calibration"},"27":{"A":"Temperature","E":"29","F":"°C","G":"Relative Humidity","K":"34","L":"% Rh"},"28":{"A":"Temperature of Calibration Liquid","E":"22.5","F":"°C","G":"Density of Calibration Liquid","K":"1.0 ± 0.002 Sp. Gr."},"29":{"A":"Uncertainty of Measurement","F":"refer next page"},"30":{"A":"\"The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k=2 such that the coverage probability corresponds to 95.45% confidence level"},"38":{"A":"DISCIPLINE : FLUID FLOW CALIBRATION"},"39":{"A":"GROUP : FLOW MEASURING DEVICES"},"40":{"A":"Calibration Result"},"42":{"A":"1.0","B":"Calibration Result of DUC \/ Error of Measurement"},"43":{"A":"Calibration is performed by considering the SI unit of Flow 1 m³\/hr = 1 m³\/hr"},"44":{"A":"Sr. No.","B":"Reading on Master","D":"Reading on DUC","F":"Error","H":"Error in % of Reading","J":"Expanded Uncertainty","L":"Coverage Factor"},"46":{"B":"( m³\/hr )","D":"( m³\/hr )","F":"( m³\/hr )","H":"( % )","J":"( % )"},"47":{"A":"1","B":"1.1081","D":"1.15","F":"0.0419","H":"3.781","J":"± 1.64","L":"2","M":"3.7812471799","N":"0.045"},"48":{"A":"2","B":"2.5418","D":"2.55","F":"0.0082","H":"0.323","J":"± 1.64","L":"2","M":"0.3226060272","N":"0.045"},"49":{"A":"3","B":"3.9908","D":"4","F":"0.0092","H":"0.231","J":"± 1.64","L":"2","M":"0.2305302195","N":"0.045"},"50":{"A":"4","B":"5.9908","D":"6","F":"0.0092","H":"0.154","J":"± 1.64","L":"2","M":"0.1535688055","N":"0.045"},"51":{"A":"5","B":"8.4908","D":"8.5","F":"0.0092","H":"0.108","J":"± 1.64","L":"2","M":"0.1083525698","N":"0.045"},"52":{"A":"→","B":"K Factor (Primary Constant) of Flow meter at the time of Calibration :","I":"0.23335"},"54":{"A":"General Remarks"},"55":{"A":"● Reporting of DUC\/Master Reading is an average of five readings ● The reading has been rounded off wherever applicable ● Calibration is performed without doing any adjustment in its original condition at time of receipt\/calibration, unless mentioned herein ● The calibration results relate only to the item calibrated and result reported in the certificate are valid at the time of and under the stated conditions of measurement. ● The Calibration certificate shall not be reproduced except in full, without written permission of the Laboratory ● *Recommended Next Due Date is reported as per requisition of customer"},"56":{"A":"-- End of Certificate --"}}},{"name":"Uncertainty","hidden":false,"max_col":"AI","rows":{"4":{"B":"Sr. No.","C":"Calibration Point","D":"Rdg on Master","E":"Rdg on Master","F":"Rdg on DUC","H":"Stdev of Master","J":"Stdev of DUC","P":"Max. Standard Deviation","Q":"UA = (Stdev\/Ön)","R":"Uncertianty of Master","S":"UB1 =(Uncertianty of Master)\/k","T":"Max. Error reported in certi","U":"UB2 = (Max. Error\/2)\/Ö3)","V":"resolution of Master","W":"UB3 = (Resol.\/2)\/Ö3)","X":"resolution of DUC","Y":"UB4 = (Resol.\/2)\/Ö3)","Z":"Density of calibration liquid (Accuracy & uncertainty of hydrometer )","AA":"UB5 = (Value)\/Ö3)","AB":"Ultrasonic Thickness Gauge (Accuracy & uncertainty of UTG )","AC":"UB5 = (Value)\/Ö3)","AD":"UC","AE":"Degree of Freedom","AF":"Coverage Factor","AG":"U","AH":"Accredited CMC","AI":"Reported Uncertainty"},"5":{"C":"m³\/hr","D":"m³\/hr","E":"m³\/hr","F":"m³\/hr","H":"m³\/hr","I":"%","J":"m³\/hr","K":"%","P":"%","Q":"%","R":"%","S":"%","T":"%","U":"%","V":"%","W":"%","X":"%","Y":"%","Z":"%","AA":"%","AB":"%","AC":"%","AD":"%","AG":"%","AH":"%","AI":"%"},"6":{"B":"1","C":"1.1","D":"1.1081","E":"1.1081","F":"1.154","G":"0.0008544004","H":"0.01177837","I":"1.0629338513","J":"0.0114017543","K":"0.988020299","L":"0.0771049882","M":"4","N":"-1","O":"3","P":"0.0771","Q":"0.0345","R":"0.22","S":"0.11","T":"1","U":"0.5774","V":"0.0090244563","W":"0.0026","X":"0.8665511265","Y":"0.2502","Z":"0.3059411708","AA":"0.1766","AB":"0.2671264786","AC":"0.1542","AD":"0.6814","AE":"608684","AF":"2","AG":"1.36","AH":"1.64","AI":"1.64"},"7":{"B":"2","C":"2.5","D":"2.5418","E":"2.5418","F":"2.551","G":"0.0001581139","H":"0.0124909968","I":"0.4914232732","J":"0.0124498996","K":"0.4880399686","L":"0.0062205478","M":"4","N":"-1","O":"3","P":"0.0062","Q":"0.0028","R":"0.22","S":"0.11","T":"1","U":"0.5774","V":"0.0039342198","W":"0.0011","X":"0.392003136","Y":"0.1132","Z":"0.3059411708","AA":"0.1766","AB":"0.2671264786","AC":"0.1542","AD":"0.6429","AE":"11117357437","AF":"2","AG":"1.29","AH":"1.64","AI":"1.64"},"8":{"B":"3","C":"4","D":"3.9908","E":"3.9908","F":"4","G":"0.0001581139","H":"0.0001581139","I":"0.0039619596","J":"0","K":"0","L":"0.0039619596","M":"4","N":"-1","O":"3","P":"0.004","Q":"0.0018","R":"0.22","S":"0.11","T":"1","U":"0.5774","V":"0.0025057633","W":"0.0007","X":"0.25","Y":"0.0722","Z":"0.3059411708","AA":"0.1766","AB":"0.2671264786","AC":"0.1542","AD":"0.6369","AE":"62698184638","AF":"2","AG":"1.27","AH":"1.64","AI":"1.64"},"9":{"B":"4","C":"6","D":"5.9908","E":"5.9908","F":"6","G":"0.0001581139","H":"0.0001581139","I":"0.0026392783","J":"0","K":"0","L":"0.0026392783","M":"4","N":"-1","O":"3","P":"0.0026","Q":"0.0012","R":"0.22","S":"0.11","T":"1","U":"0.5774","V":"0.0016692261","W":"0.0005","X":"0.1666666667","Y":"0.0481","Z":"0.3059411708","AA":"0.1766","AB":"0.2671264786","AC":"0.1542","AD":"0.6346","AE":"312849365275","AF":"2","AG":"1.27","AH":"1.64","AI":"1.64"},"10":{"B":"5","C":"8.5","D":"8.4908","E":"8.4908","F":"8.5","G":"0.0001581139","H":"0.0001581139","I":"0.0018621789","J":"0","K":"0","L":"0.0018621789","M":"4","N":"-1","O":"3","P":"0.0019","Q":"0.0008","R":"0.22","S":"0.11","T":"1","U":"0.5774","V":"0.0011777453","W":"0.0003","X":"0.1176470588","Y":"0.034","Z":"0.3059411708","AA":"0.1766","AB":"0.2671264786","AC":"0.1542","AD":"0.6337","AE":"1574834324209","AF":"2","AG":"1.27","AH":"1.64","AI":"1.64"},"11":{"C":"0","D":"0","E":"0","F":"0","G":"0","H":"#DIV\/0!","I":"#DIV\/0!","J":"#DIV\/0!","K":"#DIV\/0!","L":"#DIV\/0!","M":"4","N":"-1","O":"3","P":"#DIV\/0!","Q":"#DIV\/0!","R":"0.22","S":"0.11","T":"1","U":"0.5774","V":"#DIV\/0!","W":"#DIV\/0!","X":"#DIV\/0!","Y":"#DIV\/0!","Z":"0.3059411708","AA":"0.1766","AB":"0.2671264786","AC":"0.1542","AD":"#DIV\/0!","AE":"#DIV\/0!","AF":"#DIV\/0!","AG":"#DIV\/0!","AH":"1.64","AI":"#DIV\/0!"}}},{"name":"Software","hidden":false,"max_col":"B","rows":{"1":{"A":"Receipt No","B":"LM251224\/1"},"2":{"A":"Job No","B":"LM251224\/1 \/ TC \/ FL \/ 002"},"3":{"A":"URL No","B":"CC234525000000206F"},"4":{"A":"Issue Date","B":"25\/12\/2025"},"5":{"A":"Customer Name","B":"AIPL"},"6":{"A":"Customer Address","B":"B-95, Sector-25 GIDC, Gandhinagar, Gujarat, India, 382016"},"7":{"A":"Receipt Type","B":"Lab"},"8":{"A":"Gatepass No","B":"LTB-24122025-001"},"9":{"A":"Gatepass Date","B":"24\/12\/2025"},"10":{"A":"Receipt Date","B":"24\/12\/2025"},"11":{"A":"Calibration Date","B":"24\/12\/2025"},"13":{"A":"Due Date","B":"24\/12\/2026"},"14":{"A":"Description","B":"Digital Flow meter (Electromagnetic)"},"15":{"A":"Make","B":"Accumax"},"16":{"A":"Model","B":"AIPL_WM"},"17":{"A":"Sr No","B":"SR_19"},"18":{"A":"ID No","B":"ID_19"},"19":{"A":"Location","B":"Unit-1"},"20":{"A":"Type","B":"Digital"}}}]}